4PID Sensors (ATEX & IECEx)
Update time:Feb-02-26
Performance Characteristics Photon energy 10.6 eV Measurement range 0 ~ 10000 ppm isobutylene Resolution 1 ~ 2000 ppb isobutylene Response time (T90) 5 seconds ba seline shift (20℃) 70 15mV for detection range of0~200 0~2000 0~10000 pp
Performance Characteristics
|
Photon energy |
10.6 eV |
|
Measurement range |
0 ~ 10,000 ppm isobutylene |
|
Resolution |
1 ~ 2,000 ppb isobutylene |
|
Response time (T90) |
< 5 seconds |
|
Baseline (20℃) |
95±55mV for detection range of 0~50 ppm |
|
|
110±70mV for detection range of 0~200 ppm |
|
|
70±30mV for detection range of 0~2000 ppm,0~10000 ppm |
|
Linearity |
linear from 0.045 ~ 2.5 V |
Environmental
|
Temperature Range |
-20℃ ~ +50℃ |
|
Pressure Range |
1 atm ± 10% |
|
Humidity Range |
15% ~ 95%RH non-condensing |
Life Time
|
Storage temp |
0℃ ~ 30℃ |
|
Operating life time |
5 years (excluding lamp and electrodes) |
|
Typical lamp life |
10,000 hours |
|
Storage life |
2 years in original packaging |
|
Warranty |
12 months |
Electrical
|
Supply voltage |
3.2 ~ 5.5 V |
|
Working current |
< 35 mA at 5.0 V |
|
Output signal |
0.045 ~ 2.5 V |
Mechanical
|
Enclosure |
Stainless steel |
|
Weight |
15 grams |
Certifications IECEx: CSAE 23.0038U & Approvals
|
IECEx |
CSAE 23.0038U |
|
|
Ex ia IIC Ga |
|
ATEX |
CSANe 23ATEX1144U |
|
|
II 1G Ex ia IIC Ga |
|
|
-20℃≤Ta≤50℃ |